SURVEY RADIOMETER-DOSIMETER RKS-01G-SOLO
Description:
RKS-01G-SOLO
SURVEY RADIOMETER-DOSIMETER "RKS-01G-SOLO"
A highly sensitive survey radiometer-dosimeter RKS-01G-SOLO is designed to detect sources of gamma radiation, to arrange radiation control of the environment, work place, units and vehicles, to control scrap metal and other types of secondary materials.
The instrument is provided with a convenient survey mode for customs control and is equipped with light and sound alarm systems for radiation threshold excess values.
The instrument is entered into the RK State Register as No. KZ.02.02.00335-2004
The alignment of energy dependence non-linearity of scintillators' sensitivity is performed by means of the analog-to-digital converter.
The instrument provides for solution of tasks for customs radiation control in accordance with the requirements of RSR-99, IAEA and other regulatory documents.
PECULIARITIES:
Automatic recording of gamma background;
A programmable value of the detection threshold;
A convenient LCD and a dial surveying mode;
Assessment of measured radio nuclide spectral composition with the help of a built in analogue and digital converter, identification of radio nuclide groups;
PC-link using standard communication interface for data transmission;
Small sizes and lightweight of the instrument;
A telescopic rod.
The limit of a permissible basic relative error in measurements is: ±20%The Detector overall size:
Diameter, mm 80
Long, mm 90
Weight, kg not more than 0.3
Power supply is an internal lithium-ion battery, with a voltage of 7,2 V or through a mains adapter with an output voltage of 12 V which is included into the package
The operation range of temperatures is, oC, from -40 to +50
The control panel overall size, mm 200 х 120 х 100
The control panel weight, kg, does not exceed 1.0
The instrument is produced in 2 modifications
Modification 1
The instrument includes a control panel and gamma detector.
A gamma radiation detecting block
A scintillation detecting block
The range of gamma- radiation energies is from 20 keV to 10 MeV
The Measurement range:
of equivalent dose rate, mSv/hr 0.01-3x102
of equivalent dose, mSvr 0.1-1x108
The limit of a permissible basic relative error in measurements is: ±15%
The Detector overall size:
Diameter, mm 38
Long, mm 270
Weight, kg, not more than 0.5
Gas discharge detecting block
The range of gamma- radiation energies is from 40 keV to 10 MeV
The Measurement range:
of equivalent dose rate, mSv/hr 1х102 to 1х106
of equivalent dose, mSvr 0.1 to 1х108
The limit of a permissible basic relative error in measurements is: ±20%
The Detector overall size:
Diameter, mm 80
Long, mm 90
Weight, kg not more than 0.3
Power supply is
an internal lithium-ion battery, with a voltage of 7,2 V or through a mains adapter with an output voltage of 12 V which is included into the package
The operation range of temperatures is, oC, from -40 to +50
The control panel overall size, mm 200 х 120 х 100
The control panel weight, kg, does not exceed 1.0
Service: During the entire lifetime.
www.solo.kz e-mail: info@solo.kz, om@solo.kz